In our commitment to advocate for the highest-quality imaging standards, an important update has been made to phantom criteria and testing for 黑料网® Nuclear Medicine- and PET-accredited sites. These accreditation changes take effect Jan. 1, 2024.
黑料网 PET Accreditation Change
Effective Jan. 1, 2024, the mean background criteria for PET phantoms will be changing to 0.90-1.10 for SUV measurements.
Please refer to the article for details.
黑料网 Nuclear Medicine (NM) Accreditation Change
Effective Jan. 1, 2024, the definition of a small rectangular field camera for NM phantoms will be changing to smallest dimension < 32 cm (currently it states: longest dimension < 32 cm).
Please refer to the article for details.
Review the requirements carefully before submitting testing items for accreditation review. Requirements are subject to change at any time. Check 黑料网 requirements frequently for updates.
If you have questions, please Select "Accreditation" for Support Team and "Ultrasound/NM/PET Technical" for category.